CZF-5 horizontal and vertical combustion tester/UL94 combustion tester suitable working conditions and main technical indicators
Ambient temperature: Room temperature ~ 40℃; Relative humidity: ≤75%;
Power supply voltage and power: 220V±10% 50HZ Power 500W
Horizontal and vertical combustion test program can be switched on the touch screen, test results can be stored in the touch screen, test results can be queried;
Automatic ignition of the Bunsen burner, automatic timing, ignition time can be set at will (can be preset on the touch screen)
Bunsen burner inner diameter 9.5mm±0.5mm
Automatic ignition can be selected before the Bunsen burner is started
The Bunsen burner is automatically returned when the flame is applied, and the gas can be switched off when the Bunsen burner is returned
Gas source: LPG (plus industrial methane gas in case of arbitration);
Cabinet material: stainless steel
Cabinet thickness: 2mm
Cabinet volume: greater than 0.75m3 cubic
External dimensions: 1310*650*1560mm length, width and height (including feet) mm background black, background illumination ≤20Lux
Combustion test area volume: >0.75 cubic meters, L*S*H about 1050mm×640mm×1300mm
Instrument weight: approx. 130kg
Timing range: 0~999.9s Timing accuracy: 10s±0.2s 30s±0.2s;
Flame application time : 0~999.9S (adjustable, can be preset on the touch screen)
Afterflame time: 0 to 999.9S
Afterburning time : 0~999.9S
Linear burning rate (V) automatically calculated by PLC, displayed and saved on the touch screen
Burning angle : 0°, 20°, 45° selectable
Flame height : 10mm ~ 200mm adjustable
Flame temperature: (100-1000)℃ adjustable
Gas flow : Gas flow 105ml/min-1000ml/min adjustable
Burning lamp : 9.5±0.3mm inner diameter of the lamp, length: 100mm±10mm