WelcomeWebsite! If you have any questions, please call !
Flammability tester,burning tester,Fire Retardant Fabrics Tester_QINSUN

Hotline:

Medical Face Mask Fire Resistance Test Machine

Medical Face Mask Fire Resistance Test Machine

【Product introduction】:Medical Face Mask Fire Resistance Tester is mainly used to test the flame retardant performance of medical masks. The combustion performance of the mask is measured through a head mold wearing a mask

quote price

Product

News

Company:Qinsun Instruments Co., Ltd
Email:info@qinsun-lab.com
Phone:
Tell:
Address:

Medical Face Mask Fire Resistance Test Machine

Position: Home > Products > Others

Medical Face Mask Fire Resistance Test Machine:

 

Medical Face Mask Fire Resistance Tester is mainly used to test the flame retardant performance of medical masks. The combustion performance of the mask is measured through a head mold wearing a mask de-vice to contact with the flame at a certain linear speed.
The device is mainly used in the field of medical protection, and also can also be used for
testing flame retardant properties of  other textile coating materials.
 
Stabdard:

GB 19083-2010 
YY 0469-2011
GB 2626-2006 
BS EN136-1998 


FEATURES:

  1. PLC + touch screen control system makes the test more convenient, operated and visualized;
  2. The mask clamp is a metal human head mold, which can fully simulate the actual use state of the mask.
  3. Burner height is adjustable.
  4. Automatic timing and positioning of burner.
  5. Combustion time and flame retardant time are recorded automatically and displayed digitally
  6. Equipped with flame temperature measuring probe and flame temperature display.
  7. Automatic control system, automatic ignition
  8. The burner is equipped with LPG interface
  9. Automatic timing, automatic storage of test data


Parameters

  1. The linear velocity of automatic movement of metal head is (60 ± 5) mm / S.
  2. Flame height is adjustable 40 ± 4mm
  3. Equipped with thermocouple for flame temperature measurement and detection
  4. The flame temperature is (800 ± 50) ºC at a height of (20 ± 2) mm from the top of the burner
  5. The distance between the top of burner and the lowest point of mask is (20 ± 2) mm
  6. Burner height adjustment range is 30mm
  7. The thickness of metal head is 2-3mm which is fire resistant
  8. Timer: 0 ~ 9999.9 seconds, accuracy: ± 0.1 seconds
  9. AC220V,50Hz,50W
  10. Weight: About 30kg


Phone:+86-17740808215 Tell:+86-21-6780 0179 Address:No.2578 Minhang District Gu Dai Road, Shanghai Email:info@qinsun-lab.com sitemap